Mar 22

EBT3

EBT3

EBT3 dosimetric film is the improved version of its predecessor the EBT2. It has been specially developed to meet the needs dosimetric applications External radiation therapyEBT2 and particularly in Brachytherapy and IMRT, to meet the requirements of radiotherapy centers without processing.

further, They have improved certain aspects, such as stability between batches.

What's New

  • New look with color change.
  • Lower sensitivity to ultraviolet light.
  • A new polymer is used to adhere the layers together.
  • More stability and less change in density between batches.
  • Same active material than the previous version.
  • Symmetrical polyester layers.
  • Micro particles that eliminate Newton rings

Advantage

techniques

  • Sensitivity range of doses of 1 cGy to 800 cGy.
  • Unveiled in real time.
  • Density changes rapidly stabilized after exposure.
  • Independent dose response of energy from keV to MeV
  • Eliminates concerns caused by the scattered radiation.
  • Tissue equivalent.
  • better uniformity 1,5%
  • High spatial resolution
  • It can be managed in light of a room. There is no need for dark room.
  • Waterproof.
  • High stability against light.

economic

  • Big size
  • No need for processing
  • It does not generate chemical waste
  • Easy to cut and handle. The film may be marked with a pen and deleted if necessary.
  • competitive price compared with conventional film.

Specifications

Structure

EBT3 film is made of two sheets of active material pasted. The outer surface is polyester, which makes a robust product and allows water to be immersed.

sensitometric response

Since the absorption spectrum of the film get to the peak it is in 636 nm, sensitivity is maximized for reading red light.

Transmission densitometers commonly used with conventional film are suitable for the film EBT3. Best results are obtained red channel with a color densitometer. But nevertheless, with a filter to 636nm with a 10nm band will get the best results.

Scanners

EBT3 film can also be viewed with a film scanner or digitizer. Although many are valid, the most recommended is the brand Epson V700, it provides the best response to the film EBT3 red channel which provides more information.

Black and white scanners use a light source spectrum, which produce less response.

HeNe laser scanners (Lumisys, Molecular Dynamics) They provide the best answer to the film EBT3 because the laser has a wavelength of 633 nm, very close to the absorption peak. But nevertheless, coherent light lasers produce artifacts caused by the interaction of polarized light on the film. These artifacts can be minimized, not eliminate, using diffusers in the path of light.

energy independence

The film Radiochromic EBT3 is designed to measure absorbed dose of high-energy radiation used in the IMRT. The answer has been shown independently to energy ranges 30 KeV to MeV

Dose fractionation

The response of the film EBT3 exposed to fractions of doses, It is indistinguishable from the answer given in a single dose.

Postexposure

The film reveals EBT3 real time without processing.

Symmetry

Unlike EBT2, the film is again symmetrical EBT3.

EBT3

technical documents (PDF)